한국센서학회 학술지영문홈페이지
[ Article ]
JOURNAL OF SENSOR SCIENCE AND TECHNOLOGY - Vol. 34, No. 3, pp.198-207
ISSN: 1225-5475 (Print) 2093-7563 (Online)
Print publication date 31 May 2025
Received 28 Apr 2025 Revised 08 May 2025 Accepted 14 May 2025
DOI: https://doi.org/10.46670/JSST.2025.34.3.198

Phase-Shift Dependence of Traveling Electromagnetic Wave on Medium's Refractive Index: A Novel Concept for Simultaneous Sensing of Fluid Concentration/Composition, Temperature, and Pressure

Nader Aljabarin1, +
1Natural Resources and Chemical Engineering Department, Tafila Technical University, 66110 Tafila, Jordan

Correspondence to: + aljabarin@ttu.edu.jo

ⓒ The Korean Sensors Society
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License(https://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

This study proposes new concepts for simultaneous measurement of concentration or composition in addition to temperature, pressure, and all other intensive properties of fluid, all-in-one sensor. The concepts are based on electromagnetic principles that utilize the dependence of phase-shift of an electromagnetic wave traveling in a medium on its refractive index combined with thermodynamics. The sensor is composed of an array of multiple antennas operating at different frequencies that transmit and receive electromagnetic signals through the fluid. By measuring the phase-shift between each transmitting and receiving antenna and applying a proper refractive index model and equation of state of the fluid, the concentration/composition and all other intensive properties can be computed. In addition, this paper also proposes a simple interfacing circuit for the proposed sensor design. The proposed sensor concept was validated on a binary mixture “He-CO2” with calibration and without calibration using literature models for the refractive index and equation of state. The results of validation showed that the error in sensor measurement is negligible for calibrated sensor and about 5%–8% with literature models without calibration.

Keywords:

Fluid composition sensor, Electromagnetic measurement, Refractive index sensor, All-In-One fluid sensor

1. INTRODUCTION

1.1. Motivation

Measuring concentration/composition of fluids is very important in laboratories and industrial process pipelines for monitoring or controlling a certain process. The present work offers a new concept for the instantaneous sensing of composition/concentration in addition to temperature, pressure, and all other intensive fluid properties, if needed. These concepts utilize the principles of electromagnetics in addition to thermodynamic considerations. The unique advantage of using EM techniques over other sensing techniques lies in the speed of measurement owing to the speed of light. In addition, EM sensors need not be in direct contact with the fluid, can be clamped externally (e.g., on a pipe), can be used for small-scale (micro/nano) or large-scale (e.g., tank ), and can be used for moving or static fluids. Moreover, the sensitivity of the EM sensor depends on the working frequency, which can be adjusted as required. The EM sensor can cover the entire range of concentrations and compositions of binary or multifluid systems. The proposed EM technique can be applied to both fluids and solids; however, this study is limited to fluids only.

In previous studies, EM techniques have been successfully employed to sense the volumetric flow rate in pipes or in micro/nano fluidic conduits [1-3]. By applying the proposed concepts, all the flow process variables can be combined into one sensor for instantaneous measurement. Instantaneous measurement means a zero time constant for the sensor, which in turn adds a great advantage to process control for quick and more stable control, in addition to increasing simplicity in the design of complex control systems [4].

1.2. Literature review

Even though EM theory was developed approximately a century and a half ago [5] and the dependence of the phase shift on the refractive index is straightforward, a careful literature review reveals that this method was not used previously to measure concentration or composition of fluids. The proposed sensing concept equivalently uses two techniques which are close but not the same. These two techniques are refractometric sensor [6-8] and permittivity measurement by capacitive sensor [9-11]. The refractometric sensor measures the refractive index using an optical technique and then computes the concentration or composition from the calibrated data of concentration versus refractive index or using a preexisting concentration-refractive index model. The disadvantages of this method are that it requires a visible line of sight between the light-emitting source and the light-detecting sensor, the procedure is lengthy and can be time-consuming, and the components used are expensive and relatively large. Permittivity sensing by capacitive sensors depends on measuring the permittivity of the fluid. From this measurement the concentration can be computed using calibrated data or preexisting models for the permittivity versus concentration, or more practically, from the refractive index because the magnetic permeability is close to 1 for non-magnetic fluids [12], especially at high frequency, the usual excellent assumption is R = εr. The disadvantage of capacitive sensors is that the capacitor needs to be in direct contact with the fluid. This method is used for, or at least recommended for, small-scale applications because a large distance between the plates of the capacitor results in very low sensitivity; therefore, it is difficult or impossible to detect the variation of permittivity for compositions using the current capability of electronics [13,14]. Because the refractive index is strongly temperature- and pressure-dependent for gases, in either case of refractive index measurement, temperature and pressure are required in conjunction with the refractive index.

The objective of this study is to develop a new EM sensor that is capable of instantaneous measurement of concentration/composition in addition to temperature, pressure, and all other intensive variables, if needed. The theory of the sensor is first introduced, and then the design of the sensor with the interfacing circuit is proposed; the concepts are validated experimentally, and conclusions are drawn.


2. THEORY

2.1. Electromagnetic considerations

When an EM wave travels in a medium, its behavior is governed by Maxwell's four Eqs. [15]:

·D = ρv(1) 
·B = 0 (2) 
 X E = -Bt(3) 
 X H = J +Dt(4) 

Considering one dimensional EM wave traveling with velocity u in charge free (ρv = 0) medium in the z-direction, the solution of Maxwell's equations in Cartesian coordinates is [16]:

Ez,t= E0e-azcosωt - βzax (5) 

and

Hz,t = E0ηe-azcosωt - βz-θηay(6) 

where, α is the attenuation constant of the medium and β is the phase constant given by

α = ω με21 + σωε2- 1(7) 
β = ω με21 + σωε2+ 1(8) 

η is the intrinsic impedance given by

η=μ/ε1 + σωε21/4(9) 

for 0 ≤ θη ≤ 45° where

tan2θη =σωε(10) 

The EM wave traveling velocity is given by

u =1με(11) 

For non-conducting fluid σ = 0 or for fluid with σ << ωε Eqs. (7) and (8) become α = 0, β = ωμε; η= μ/εand θη = 0. Therefore, Eqs. (5) and (6) become

Ez,t=E0cosωt-ωμεzax(12) 
Hz,t=E0μ/εcosωt-ωμεzay(13) 

The refractive index, R, of a medium is defined as the ratio of speed of light in vacuum, C=1μ0ε0=299792458 m/s, to the speed of light in the medium, that is

R=Cu=Cμε=1μ0ε0με=μrεr(14) 

Therefore, Eqs. (12) and (13) become

Ez,t=E0cosωt-ωRczax=E0θax(15) 
Hz,t=E0μ/εcosωt/ωRcZay=E0μ/εθay(16) 

or

Hz,t=ε4cE0Rcosωt-ωRczay=ε4cE0Rθay(17) 

where θ=ωRcz is the phase shift of the EM wave. Therefore, by sending EM wave through fluid at a fixed certain point and receiving it at a second fixed point, the phase shift of the wave at the second point will shift by ωRcz depending on the refractive index of the fluid. The refractive index then can be calculated from the measured phase shift θ as

R=cθωz(18) 

In addition, the amplitude of the magnetic field also depends on the refractive index, H=ε4cE0R. This concept can also be utilized to detect R; however, because the amplitude is proportional to ε4 and ε is in the order of 10-12 F/m, it is impossible, or at least impossible, to accurately detect the amplitude. Therefore, the only practical choice for detecting R is phase shift. The sensitivity of the sensor from Eq. (18) can be given as

ΔθΔR=ωzc,(19) 

which is linear with frequency; therefore, the sensitivity can be increased as required by increasing the frequency. This is an excellent advantage in selecting the phase shift as a measure of the refractive index.

2.2. Thermodynamic considerations

This section discusses a mathematical model to obtain the concentration/composition, temperature, pressure, and all other properties, if needed, from the calibration data or a mathematical model of the dependence of the refractive index on the concentration/composition, temperature, pressure, and frequency. According to Gibbs phase rule F = 2 - p + n, where F is the number of degrees of freedom, p is the number of phases, and n is the number of chemical species present in the mixture. The refractive index can be specified by two independent intensive properties in addition to the composition xi. In addition, it has been experimentally proven that the refractive index is a function of frequency [17-19] (Frequency is not an intensive property; therefore, the Gibbs rule is not violated). Thus, the general case for the refractive index can be selected as

R = RT, R, xi, ω, i = 1, , n-1(20) 

Because i=1nxi=1, only n-1 of mole or mass fraction needs to be specified. Differentiating both sides of Eq. (20) yields

dR=RTP,x,ωdT+RPT,x,ωdP+RxiP,T,ω,xj,jidxi+RωP,T,xdω(21) 

where xj, ji indicates that all mole fractions from j=1,...,n-1 except component i are held constant. Referring to Eq. (21), we obtain

RωT,x=RPT,x,ωPωT,x+RωP,T,x(22) 

From which the pressure can be obtained as:

PωT,x=RωT,x-RωP,T,xRPT,x,ω(23) 

or

P=P0+ω0ωRωT,x-RωP,T,xRPT,x,ωdω Constant T and x(24) 

From Eq. (18)

RωT,x=cz1ωθωT,x-θω2(25) 
RωP,T,x=cz1ωθωP,T,x-θω2(26) 
RPT,x,ω=cz1ωRPT,x,ω(27) 

Substituting (25), (26), and (27) into (24) yields

P=P0+ω0ωθωT,x-θωP,T,xθPT,x,ωdω Constant T and x(28) 

Therefore, by sending EM waves at different frequencies and measuring the refractive index at these frequencies, θωT,x is measured by the sensor, and θωP,T,x and θPT,x,ω are obtained from the calibration data/models. Hence, the pressure is obtained from Eq. (24) or (28). From Eq. (21), we obtain

RωP,x=RTP,x,ωTωP,x+RωP,T,x(29) 

In a similar step as above, the temperature can be calculated as

T=T0+ω0ωθωP,x-θωP,T,xθTP,x,ωdω Constant P and x(30) 

Also, from Eq. (21):

RωP,T= i=1n-1RxiP,T,ω,xj,jixiωP,T+ RωP,T,x(31) 

or

i=1n-1RxiP,T,ω,xj,jixiωP,T= RωP,T,x- RωP,T(32) 

Therefore, the composition can be determined using Eq. (32). For example, for binary system Eq. (32) reduces to:

Rx1P,T,ωx1ωP,T= RωP,T,x- RωP,T(33) 

Then

x1=x1,0+ω0ωRωP,T,x-RωP,TRx1P,T,ωdωConstant P and T.(34) 

By using Eq. (18) we obtain

x1=x1,0+ω0ωθωP,T,x-θωP,Tθx1P,T,ωdωConstant P and T(35) 

and x2 = 1 - x1. T0, P0, x1,0, ω0 all are the values at the base point (center point of calibration). Once the temperature, pressure, and composition are known, the refractive index can be determined from the calibration data/model. All other intensive properties (density, viscosity, enthalpy, etc.), if required, can be computed based on the thermodynamic Gibbs phase rule. This can be achieved using equations of state and/or empirical correlations. Aspen HYSYS software [20] is an excellent tool for this purpose.

2.3. Design considerations

2.3.1. Proposed design of transmitting and receiving antennas

Although many antenna designs have been proposed to transmit and receive EM signals through a fluid to measure the refractive index [21-23], in this study, a simple and practical design is proposed. Fig. 1 shows the proposed design and the circuit used to transmit and receive signals. Circular antennas are selected because they are suitable for circular pipes, particularly for clamping-on configuration. Because EM waves travel through a fluid with the speed of light, which is much higher than the fluid velocity, it makes no difference whether the fluid is static or in motion when a sensor is used. Therefore, the sensor can be used for flow streams or samples collected in tubes for laboratory measurements.

Fig. 1.

Suggested interfacing circuit for each single transmitting/receiving antennas.

The resistance 10 KΩ is required to close the circuit for electricity to flow in the circuit, and simultaneously its value should be much higher than the radiation resistance of the antenna to force the electricity to transmit through the antenna and not through the resistance itself. The order of magnitude of the resistance was estimated by comparing the proposed antenna with a Hertzian dipole antenna. The radiation resistance of Hertzian dipole is in the order of 2 Ω [16]; therefore, the resistance should be much larger than 2 Ω (10 kΩ is believed to be satisfactory for the proposed design). To measure the composition/concentration, temperature, and pressure simultaneously, multitransmitting and corresponding receiving antennas (Fig. 1) must be used. The number of these antennas should be one for temperature, one for pressure, and n - 1 for composition, for a total of 1 + 1 + n - 1 = n + 1 transmitting/receiving antennas (array). Fig. 2 shows the final design of the EM array of the sensor.

Fig. 2.

The proposed sensor design. The number of transmitting/receiving antenna arrays must be (n + 1).

2.3.2. Electromagnetic shielding

The purpose of EM shielding is to prevent the interaction/interference of atmospheric signals (communication and probably other signals) with sensor signals. The atmosphere is full of signals that can be captured by the transmitting/receiving antennas of the sensor and hence disturb or at least create noise in the signal. Although the interfering signals can be filtered in the circuit design, the use of EM shielding is much better, cheaper, and simpler. EM shielding can be performed by wrapping the sensor with an insulation material and then wrapping it with a conductive metal (e.g., copper or aluminum). The function of the insulation material is to prevent contact between the transmitting/receiving antennas and shield the metal. The thickness of the metal shield can be estimated by calculating the skin depth given by [16]

δ =1πfμσ.(36) 

The metal thickness of five times the skin depth was sufficient to block the interfering signals almost completely. Practically, based on Eq. (36), the thickness of the metal shielding is on the order of fractions of one-millimeter. Therefore, a metal thickness of 1 mm is sufficient for EM shielding.

2.3.3. Limitations

The proposed sensor design has the following limitations:

i. If the pipe/tube is a metal, the transmitting/receiving antennas must be installed internally inside the pipe with an insulation layer between the antennas and the pipe. Metals absorb and therefore prevent EM signals from traveling through the walls of the pipe if they are made of metal (Eq. (36)).

ii. The proposed concept cannot be used to detect the pressure if the fluid under measurement is a liquid. The refractive index is weakly dependent on the pressure of the liquid [24]. Hence, RPT,x,ω in Eq. (24) or θTP,x,ω in Eq. (28) are zeros and these equations cannot be used to calculate the pressure. However, the concentration, temperature, and all other intensive properties can still be measured in the same manner as explained above.


3. EXPERIMENTAL VALIDATION OF THEORETICAL CONCEPTS

3.1. Experimental setup

Fig. 3 shows the equipment used to validate the proposed concepts and sensor design. Binary Helium (He) and carbon dioxide (CO2) gases were used for validation. Two cylinders with relatively high pressure of He and CO2 are connected to a glass cylinder (glass was selected for visibility). A thermometer and pressure gauge were placed inside the glass cylinder to measure the temperature and pressure, respectively. A stirrer was placed inside the glass cylinder to ensure a good mixing of the two gases. The temperature and pressure inside the glass cylinder were controlled using an electrical heater and hand valves, respectively. A sinusoidal signal was generated using a functional signal generator, and the phase shift was measured using an oscilloscope.

Fig. 3.

Equipment used for sensor validation: 1. Sensor, 2. Glass Cylinder, 3. Heater, 4. Electrical Motor, 5. Circumferential Bolts, 6. Pipes, 7. Pressure Gauges, 8. Stirrer, 9. Hand Valves, 10. He Cylinder, 11. CO2 Cylinder 12. Thermometer, 13. Signal Functional Generator, 14. Oscilloscope.

At the beginning of the experiment, the glass cylinder was flushed with He gas to expel air, and the He pressure was set to the required level. Subsequently, the He valve was closed so that its amount remained constant for the next course of the experiment. The specific molar volume of He inside the glass cylinder was calculated using the measured temperature and pressure. After the He valve was closed, the CO2 valve was opened, and the amount of CO2 inside the glass cylinder was increased to the required level. The temperature of the gas mixture inside the glass cylinder was controlled by turning on the heater and increasing or decreasing the input power to obtain the required temperature. At each point, the temperature and pressure were recorded, and the mole fraction of CO2 was calculated as follows:

xco2=nco2nco2+nHe=nt-nHent-nHe+nHe=nt-nHent=1-nHent(37) 

Assuming ideal gas we obtain

xco2=1-nHent=1-vHevt=1-PiRTiPRT=1-TTiPiP(38) 

or

xco2=1-TTiPiP(39) 

After estimating the mole fraction of CO2 from Eq. (39), the CO2 mole fraction was recalculated using the Peng-Robinson equation of state [25] with the help of the Aspen HYSYS software [20]. Owing to the initial He content, the mole fraction of CO2 did not reach one. Therefore, to obtain a mole fraction of one for CO2, the experiment was repeated; however, in this case, the experiment was started with CO2 and He was subsequently added. Therefore, two sets of experiments were conducted: in the first set, He was initially present and CO2 was then added to the He, whereas in the second set, CO2 was initially present and He was added to the CO2. Thus, all mole fractions from 0–1 were covered.

3.2. Validation based on calibration

The equipment shown in Fig. 3 was used to calibrate the sensor over the ranges listed in Table 1. The phase shift was measured at different temperatures, pressures, mole fractions, and frequencies, and Eqs. (28), (30), and (35) were used for calibration.

Calibration range of the sensor based on equipment capability.

3.3. Validation based on literature models of refractive index

If a reliable model for the refractive index is available, R = R(T,P,xi,ω)then the sensor can be used without calibration. The proposed sensor was validated based on separate literature models of the refractive indices. For He, the following model of refractive index as a function of wavelength was selected [26]:

nHe- 1 =0.001470091423.98 - λ-2(40) 

For CO2, the dependence of the refractive index on the wavelength was selected as [27]:

NCO2-1=0.000001544890.0584738-λ-2+0.083091927210.9341-λ-2+0.002876419060.122959-λ-2(41) 

where λ in Eqs. (40) and (41) is the vacuum wavelength (C = λω) and the conditions of the equations are: T = 0°C and P = 101325 Pa. The Lorentz–Lorenz equation [28] was selected to determine the dependence of the refractive index on density.

R2- 1R2+ 2= κρ(42) 

Assuming (κρ) = xHev(κρ)He + xCO2(κρ)CO2, we obtain Lorentz–Lorenz mixing rule as

R2-1R2+2=xHeRHe2-1RHe2+2+xCo2RCO22-1RCO22+2(43) 

Because the Lorentz–Lorenz equation is dependent on density rather than temperature and pressure, the Peng-Robinson equation of state with the help of the Aspen HYSYS software was selected to relate the density with pressure and temperature. Eqs. (24), (29), and (34) are used in this case.

3.4. Results of validation

Fig. 4 shows the representative results of pressure versus CO2 mole fraction at a fixed temperature for both sets of experiments, as discussed in Section 3.1. The figure shows excellent agreement between the measured values and the calibrated sensor measurement and good agreement for the sensor measurement without calibration when the literature models were used. The figure also shows that for the literature models, the sensor measurement is underpredicted. Fig. 5 shows a temperature versus pressure plot for a fixed CO2 mole fraction. Again, this figure shows excellent agreement for the calibrated sensor and good agreement for the sensor measurements based on the literature models. In this case, the temperature was overpredicted.

Fig. 4.

Experimental results obtained for pressure versus CO2 mole fraction at 40.1oC.

Fig. 5.

Experimental results obtained for temperature as a function of pressure at CO2 mole fraction of 0.48.

Finally, Fig. 6 shows a bubble plot for random points of T, P, and mole fractions. Again, the figure shows excellent agreement for the calibrated sensor and good agreement for the model sensors reported in literature. The error in the sensor measurements was calculated as

Fig. 6.

Bubble plots of random points for P, T and CO2 mole fraction. The size of bubbles indicate the CO2 mole fraction.

Error=1N1N|Experimental Value-Sensor Measured VlaueExperimental Value|×100%.(44) 

Table 2 lists the errors based on all the validation points. The table shows negligible errors for the calibrated sensor and acceptable errors for the sensor when the literature models were used.

Error percentage of sensor measurement.


4. CONCLUSIONS

In this study, a novel sensor for measuring the concentration/composition, temperature, pressure, and all other intensive properties of fluids was successfully developed. The sensor can be used without calibration if reliable models for the refractive index are available in the literature; otherwise, calibration is required. The sensor concepts were validated using a "He-CO2"system, and their accuracy was verified. With calibration the sensor errors were less than 1%, whereas without calibration using literature models of refractive index the errors were 7% for the mole fraction, 5% for temperature, and 8% for pressure.

Because a general refractive-index mathematical model that depends on temperature, pressure, composition, and frequency is usually not available in the literature, a relatively simple procedure for calibrating the sensor and/or using separate literature models was developed based on thermodynamic principles. This calibration procedure, although mathematically exact, allows decoupling of the variables and makes the calibration for each variable independent, which significantly simplifies the procedure. The procedure was experimentally validated.

To use the proposed sensors or concepts proposed in this work, it is recommended that the refractive indices of the materials be significantly different. The sensor can still be used for materials with slightly different refractive indices; however, this requires precise calibration and the design and fabrication of electronic circuits, which can be difficult to achieve. The as-developed sensor fails completely if the refractive indices of the materials are the same. Of course, this is practically impossible.

Glossary

Nomenclature
D : Electric flux density, C/m2
B : Magnetic flux density, Wb
E : Electrical field strength, V/m
H : Magnetic field intensity, A/m
J : Current Flux, A/m2
C : Speed of light in vacuum, m/s
t : time, s
ρv : Charge density, C/m3
ω : Frequency, rad/s
μ : Permeability, H/m
ε : Permittivity, F/m
σ : Electrical conductivity, S/m
θ : Phase shift, rad
u : Velocity, m/s
c : Speed of light in medium, m/s
R : Refractive index
F : Number of degrees of freedom
p : Number of phases
n : Number of chemical species present in the mixture
T : Temperature, oC or K
P : Pressure, Pa
x : Mole or mass fraction
ƒ : Frequency, Hz
δ : Skin-depth, m
n : Number of species, mole
λ : Wavelength, m
N : Number of experimental points
ρ : Density, Kg/m3
v : Specific volume, mole/m3
Subscripts
v : Volume
t : Total
i : Initial

Acknowledgments

The author thanks the financial support provided by Tafila Technical University and is grateful to the editor and referees for carefully reading the paper and for their comments and suggestions, which have improved it.

References

  • A.M. Al Jarrah, N. Aljabarin, Magnetic Coupling of Two Coils Due to Flow of Pure Water Inside Them–Double Coil Volumetric Flow Sensor, Adv. Sci. Technol. Res. J. 16 (2022) 47–53. [https://doi.org/10.12913/22998624/147973]
  • A.M. Al Jarrah, A Novel Simple Technique for Measuring the Volumetric Flow Rate and Direction of Flow Inside a Pipe–The Single and Double Coils Sensors, Adv. Sci. Technol. Res. J. 16 (2022), 290–298. [https://doi.org/10.12913/22998624/155038]
  • A.M.A. Jarrah, M.M. AlMahasneh, Using Parallel Plates Capacitor as A Volumetric Flow Rate Sensor and Direction Detection for Microfluidic/Nanofluidic and Extra Smaller Applications, Sens. Int. 4 (2023) 100247. [https://doi.org/10.1016/j.sintl.2023.100247]
  • C.A. Smith, A.B. Corripio, Principles and Practices of Automatic Process Control, 3rd ed., Wiley, Hoboken, 2005.
  • A Treatise on Electricity and Magnetism. https://en.wiki­pedia.org/wiki/A_Treatise_on_Electricity_and_Magnetism, (accessed April 2025).
  • S. Pissadakis, S. Selleri, Optofluidics, Sensors and Actuators in Microstructured Optical Fibers, Woodhead Publishing, Cambridge, 2015.
  • J.L. Santos, F. Farahi, Handbook of Optical Sensors, CRC Press, CRC Press, Boca Raton, 2015.
  • J. Haus, Optical Sensors: Basics and Applications, Wiley-VCH, Weinheim, 2010. [https://doi.org/10.1002/9783527629435]
  • L.K. Baxter, Capacitive Sensors: Design and Applications, Wiley-IEEE Press, Piscataway, 1997.
  • W.Y. Du, Resistive, Capacitive, Inductive, and Magnetic Sensor Technologies, CRC Press, Boca Raton, 2015.
  • J.R.S. Avila, K.Y. How, M. Lu, W. Yin, A Novel Dual Modality Sensor with Sensitivities to Permittivity, Conductivity, and Permeability, IEEE Sens. J. 18 (2017) 356–362. [https://doi.org/10.1109/JSEN.2017.2767380]
  • Refractive index. https://en.wikipedia.org/wiki/Refractive_index, (accessed April 2025).
  • W.G. Jung, Op Amp Applications Handbook, Newnes, Burlington, 2005. [https://doi.org/10.1016/B978-075067844-5/50152-1]
  • P. Scherz, S. Monk, Practical Electronics for Inventors, McGraw-Hill Education, New York, 2016.
  • D.K. Cheng, Field and Wave Electromagnetics, 2nd ed., Pearson Education, Harlow, 1989.
  • M.N.O. Sadiku, S. Nelatury, Elements of Electromagnetics, 7th Edition, Oxford University Press, New York, 2018.
  • D. Malacara, B.J. Thompson, Handbook of Optical Engineering, CRC Press, Boca Raton, 2001. [https://doi.org/10.1201/9780203908266]
  • R. Kingslake, Applied Optics and Optical Engineering, 1st Ed., Academic Press, Cambridge, 2012.
  • T.P. Otanicar, P.E. Phelan, J.S. Golden, Optical Properties of Liquids for Direct Absorption Solar Thermal Energy Systems, Sol. Energy 83 (2009) 969–977. [https://doi.org/10.1016/j.solener.2008.12.009]
  • Aspen Technology Inc., Aspen HYSYS Software V14.
  • Y.T. Lo, S.W. Lee, Antenna Handbook: Theory, Applications, and Design, Springer Science & Business Media, New York, 2013.
  • C.A. Balanis, Modern Antenna Handbook, Wiley, Somerset, 2011.
  • D.K. Barton, S.A. Leonov, Radar Technology Encyclopedia, Artech House, Boston, 1997.
  • M. Omini, Optical Properties of Liquids Under Pressure, J. Phys. 39 (1978) 847–861. [https://doi.org/10.1051/jphys:01978003908084700]
  • J. Gmehling, M. Kleiber, B. Kolbe, J. Rarey, Chemical Thermodynamics for Process Simulation, 2nd, Completely Revised and Enlarged Edition, Wiley-VCH, Weinheim, 2019. [https://doi.org/10.1002/9783527809479]
  • C. R. Mansfield, E. R. Peck, Dispersion of Helium, J. Opt. Soc. Am. 59 (1969) 199–204. [https://doi.org/10.1364/JOSA.59.000199]
  • J.G. Old, K.L. Gentili, E.R. Peck, Dispersion of Carbon Dioxide, J. Opt. Soc. Am. 61 (1971) 89–90. [https://doi.org/10.1364/JOSA.61.000089]
  • H. Kragh, The Lorenz-Lorentz Formula: Origin and Early History, Substantia 2 (2018) 7–18.

Fig. 1.

Fig. 1.
Suggested interfacing circuit for each single transmitting/receiving antennas.

Fig. 2.

Fig. 2.
The proposed sensor design. The number of transmitting/receiving antenna arrays must be (n + 1).

Fig. 3.

Fig. 3.
Equipment used for sensor validation: 1. Sensor, 2. Glass Cylinder, 3. Heater, 4. Electrical Motor, 5. Circumferential Bolts, 6. Pipes, 7. Pressure Gauges, 8. Stirrer, 9. Hand Valves, 10. He Cylinder, 11. CO2 Cylinder 12. Thermometer, 13. Signal Functional Generator, 14. Oscilloscope.

Fig. 4.

Fig. 4.
Experimental results obtained for pressure versus CO2 mole fraction at 40.1oC.

Fig. 5.

Fig. 5.
Experimental results obtained for temperature as a function of pressure at CO2 mole fraction of 0.48.

Fig. 6.

Fig. 6.
Bubble plots of random points for P, T and CO2 mole fraction. The size of bubbles indicate the CO2 mole fraction.

Table 1.

Calibration range of the sensor based on equipment capability.

Variable Measurement Accuracy (±) Range of Calibration
Temperature, oC 0.05 22.0–55.0
Pressure, kPa 0.5 106.4–314.1
CO2 Mole Fraction 0.01 0–1.00
Frequency 0.005 Hz 0 –100 MHz

Table 2.

Error percentage of sensor measurement.

Variable Calibrated Sensor (%) Sensor Measurement Based on Literature Models of Refractive Index (%)
CO2 mole fraction Negligible (< 1%) 7
Temperature Negligible (< 1%) 5
Pressure Negligible (< 1%) 8